(12) When oil-free water clarity changes, this change can be calibrated out by
running a sample of the oil-free water through unit and adjust zero
adjustment knob (10) to obtain reading of 00 on digital display (4).
NOTE
Release knob lock before attempting to turn
knob.
(13) Depress CAL check switch (11) and reestablish calibration check number
(5).
(14) Close side panel of control module.
2-235
|
||
![]() |